Regulatory 9 min read

How to Write a PSUR Under EU MDR: A Step-by-Step Guide

Who needs a periodic safety update report, when it is due, the sections MDCG 2022-21 recommends, and how to present the data and the benefit-risk conclusion.

DR
Data Reality Team
How to Write a PSUR Under EU MDR: A Step-by-Step Guide

A PSUR (periodic safety update report) is the report that Article 86 of the EU Medical Device Regulation (MDR) requires for class IIa, IIb and III devices: a summary of the results and conclusions of your post-market surveillance data, with the benefit-risk conclusion, the main findings of post-market clinical follow-up (PMCF), and the sales and patient population figures for the period. To write one, you collect the data your post-market surveillance plan names for the data collection period, present it the way the MDCG 2022-21 guidance recommends, compare it with earlier periods, and end with a clear, evidence-based statement on whether the device's benefit-risk profile has changed.

This guide walks through each step: which devices need a PSUR and how often, when the data collection period starts, how to group devices, the sections to include, how to present the numbers, and how to write the conclusion. It follows the text of the MDR and of MDCG 2022-21, the Medical Device Coordination Group's guidance on the PSUR (December 2022). The guidance is not legally binding, but it is the MDCG's explanation of how to meet Article 86.

What are the PSUR requirements under EU MDR?

Article 86 of the MDR asks manufacturers of class IIa, IIb and III devices to prepare a PSUR for each device and, where relevant, for each category or group of devices. The PSUR summarises the results and conclusions of the analyses of the post-market surveillance data gathered under the post-market surveillance plan, together with a rationale and description of any preventive and corrective actions taken.

Throughout the lifetime of the device, the PSUR must set out three things:

  • The conclusions of the benefit-risk determination.
  • The main findings of the PMCF.
  • The volume of sales and an estimate of the size and other characteristics of the population using the device and, where practicable, how often it is used.

The PSUR is part of the technical documentation on post-market surveillance described in Annexes II and III of the MDR (for custom-made devices, part of the documentation in Annex XIII). MDCG 2022-21 adds that it should be a stand-alone document that a reviewer can assess without opening the supporting reports: it summarises the results and conclusions of your post-market surveillance activities rather than copying every dataset.

Class I devices don't need a PSUR. Their manufacturers prepare the simpler post-market surveillance report (PMSR) under Article 85, updated when necessary and given to the competent authority on request.

Which devices need a PSUR, and how often?

The device class sets how often the PSUR is updated and where it goes. Class III and implantable devices are submitted through EUDAMED; the others are kept in the technical documentation and made available to the notified body.

DeviceReportUpdate at leastWhere it goes
Class III, and class IIb implantablePSUREvery yearSubmitted to the notified body through EUDAMED; the notified body adds its evaluation
Class IIa implantablePSUREvery two yearsSubmitted to the notified body through EUDAMED; the notified body adds its evaluation
Class IIb, not implantablePSUREvery yearMade available to the notified body, and to competent authorities on request
Class IIa, not implantablePSURWhen necessary, and at least every two yearsMade available to the notified body, and to competent authorities on request
Class IPMSR (Article 85)When necessaryKept in the technical documentation; given to the competent authority on request

Until EUDAMED is in use for PSURs, manufacturers of class III and implantable devices deliver the PSUR to their notified body by a method agreed with it (MDCG 2022-21, section 5.4). Custom-made devices of class IIa, IIb and III also need a PSUR on the same schedule, but it stays in the custom-made device documentation and is not submitted through EUDAMED.

The obligation lasts for the lifetime of the device: a PSUR is needed until the last device placed on the market has reached the end of its intended lifetime, so it can continue for years after the product is discontinued. If the lifetime is longer than the documentation retention period (10 years, or 15 for implantable devices), the shorter of the two applies (MDCG 2022-21, section 3.2).

When does the PSUR data collection period start?

For a device certified under the MDR, the data collection period starts on the device certification date. MDCG 2022-21 (section 1.2) defines it as the date of the EU technical documentation assessment certificate, the EU type-examination certificate or the EU product verification certificate; for devices assessed under Annex IX Chapters I and III, the date the declaration of conformity was signed; and for custom-made devices, the date of the Annex XIII statement. Each period ends on the anniversary of that MDR certificate, unless the manufacturer agrees another schedule with its notified body.

  • The first PSUR is due one year (class IIb and III) or two years (class IIa) after the certification date. A newly certified device does not need a PSUR at the moment of certification.
  • Periods are contiguous. The end of one data collection period is the start of the next, with no gap or overlap. The first PSUR may therefore cover a period that is not exactly 12 or 24 months.
  • Legacy devices (still sold under an MDD or AIMDD certificate during the MDR transition) follow the class of their MDD certificate, start their data collection on 26 May 2021, and are made available on request rather than submitted through EUDAMED.

Can one PSUR cover several devices?

Yes. A PSUR can cover one device, meaning one Basic UDI-DI with its variants and sizes, or a group of devices, but you must justify the grouping. MDCG 2022-21 sets these rules for grouped PSURs:

  • Choose a leading device. It must be the highest risk class in the group (or one of them), and it sets the schedule for the whole PSUR: the data collection period, the update frequency and whether it goes through EUDAMED.
  • Keep the leading device. Other devices can be added or removed in later updates, but the leading device cannot change. A new leading device means a new PSUR.
  • Use one notified body. Only devices assessed by the same notified body can share a PSUR.
  • Show each device separately. Present the data so a reader can see how each Basic UDI-DI performs on its own.

What sections should a PSUR include?

Annex I of MDCG 2022-21 gives a recommended template, and the guidance suggests using the same presentation whatever the device class.

SectionWhat goes in it
Cover page and table of contentsManufacturer, devices covered, notified body name and number, PSUR reference number, version number and the data collection period
Executive summary (recommended)Status of actions from the previous PSUR, by you and by the notified body; any change to the data collection period; a clear statement on the benefit-risk profile
Description of the devices and intended useRisk class, first certification or market date, current status, intended purpose, indications and target populations, by Basic UDI-DI with trade names and EMDN codes
Grouping of the devicesWhy the devices share one PSUR, and which is the leading device
Volume of salesDevices placed on the market, year by year, using one consistent measure
Size and characteristics of the populationEstimated number of patients exposed, usage frequency, and relevant demographics
Vigilance and CAPASerious incidents, trend reports, field safety corrective actions and safety-related preventive and corrective actions
General PMCF informationComplaints and user feedback, literature review, registries and public databases, information about similar devices
Specific PMCF informationMain findings of PMCF studies and other specific PMCF activities
Findings and conclusionsValidity and limits of the data, overall conclusions on benefits and risks, and actions taken

If a dataset from your post-market surveillance plan isn't used, say so in the relevant section and justify why: MDCG 2022-21 asks manufacturers to justify any dataset they leave out.

How should you present the data in a PSUR?

Annex III of MDCG 2022-21 describes how the data should be presented. The main points:

  • Compare with earlier periods. Each PSUR shows the current period next to the same data from previous periods. For class IIb and III, that means the latest 12 months and the three 12-month periods before it; for class IIa, the latest 24 months and the 24 months before.
  • Split by region. Where applicable, show the EEA, Turkey (TR) and Northern Ireland (XI) separately from worldwide data, which includes them.
  • Split by device. Present each dataset by Basic UDI-DI, or by model if there is no Basic UDI-DI.
  • Give rates as well as counts. Report serious incidents as absolute numbers and as rates where applicable, and give complaint occurrence rates with the reference you used, such as units sold.
  • Use IMDRF terms where available. Report serious incidents with the IMDRF Adverse Event Terminology, for example device problems (Annex A) and health impacts (Annex F); Level 2 terms are enough for grouping. Complaints can be grouped by IMDRF device problem codes or by your internal event codes, with their terms.
  • Keep the sales measure consistent. Units sold, units shipped, implants or episodes of use: pick one, state it, and use it everywhere, because it is the denominator of your rates.

Annex II of the guidance has example tables for sales, serious incidents, field safety corrective actions and CAPA.

How do you write the PSUR benefit-risk conclusion?

The conclusion is where the PSUR answers the question Article 86 puts first: what the data means for the benefit-risk determination. It is written by your team. MDCG 2022-21 asks for three things:

  • Validity of the data. Name the limits of what you collected, such as low sales, known bias in feedback or slow PMCF enrolment, and say whether they still allow a meaningful conclusion.
  • Overall conclusions. Describe any new or emerging risks, poor performance or benefits not achieved, including which patient groups, models or regions they affect and how serious they are. Describe new benefits too. Then state, based on the evidence, whether the benefit-risk profile has been adversely affected or remains unchanged.
  • Actions taken. List the actions started during the data collection period to address new risks or poor performance.

MDCG 2022-21 recommends an executive summary; when you include one, it should state the bottom line in one clear statement, for example that the benefit-risk profile of the device has not been adversely impacted and remains unchanged. Avoid hedged wording: the reader should not have to infer your conclusion from the tables.

What are the most common PSUR mistakes?

  • Counts without rates. Ten complaints mean little without the number of devices sold in the same period.
  • No comparison with previous periods, so the reader can't see whether things are getting better or worse.
  • Missing regional split between the EEA, Turkey, Northern Ireland and worldwide data.
  • Summary text that disagrees with the tables after a late data correction.
  • Datasets left out without a justification.
  • A conclusion that lists data but doesn't judge it. The PSUR has to say what the data means for the benefit-risk profile.

How can PSUR software help?

Much of the work in a PSUR is counting: sales by region and year, complaint and incident rates, and the same tables again for every earlier period and every device group. That part can be calculated. With PSUR Wizard you build the PSUR in your own template: tables, charts and the numbers in every summary are calculated from your sales, complaint and vigilance data, and when a data file is corrected and uploaded again, the whole report recalculates. Your team reviews the document, writes the assessment and the benefit-risk conclusion, and downloads the finished PSUR in Word.

Where can you read the PSUR rules in full?